3.4
Removing the Monitor Unit
hAUTION:
T wo persons
are required
when removing
the monitor
unit.
If a worker
removes
the monitor
alone, he/she may be injured.
3.4.1
Removing the monitor unit (for systems with the CRT monitor)
(1)
Removing the monitor rear cover
Loosen the two screws CA> and four screws cB> that secure the monitor rear cover to
remove the cover.
(2)
Removing the support rear cover
The support rear cover engages with the support front cover.
Disengage the upper part of the cover and remove the cover by pulling it backward.
(3)
Removing the cables (refer to the next page)
(a)
Loosen the two screws CC> to remove the pinch shield cover.
(b)
Remove the two screws CD> to remove the clamps.
(c)
Remove the signal, audio, and power cables that are connected to the monitor unit.
(4)
Removing the monitor unit from the system
Remove the four M6 bolts <E> to remove the monitor unit with the boss installed.
-
Screw <A> (two)
I
Support rear cover
3-12