OPERATION
3.7
Output Set-Up
The source and scale of the 4-20 mA output can be confi gured in
settings sub menu.
From the Main Reading Page, press
to scroll to the
The top line shows the navigation command. Press the key to change the navigation
command from
Press the
Select
Reading Page or enter the output settings sub-menu. The fi rst setting parameter under
this sub-menu is the output source selection.
3.7.1
Output Source Selection
The following list shows the options for the output source:
PPMV_N – moisture content in ppm
MG/M3 – moisture content in mg/m
LBMMCF – moisture content in lb/MMscf natural gas
PPMV_I – moisture content in ppm
DP_F – °F dew point
psiG – pressure in psig
DP_C – °C dew point
BARG – pressure in barg
a.
Change the selection by scrolling through these options using the and
keys.
b.
When the desired source is on the display, press the
The 4-20 mA output will then switch to the selected source. The display
will change to the
40
Menu
option.
OUTPUT
O U T P U T
to
.
SKIP
ENTER
O U T P U T
key to confi rm the navigation command and either skip to the Main
S O U R C E
V
value for the output range set-point.
ZERO
Promet I.S. User's Manual
to enter the main menu. Use the
S K I P
E N T E R
M O I S T
natural gas
V
natural gas
3
ideal gas
Select
– output
OUTPUT
Select
key to confi rm.
97221 Issue 5, March 2017
key