3.8.3

Alarm Source Selection

The alarm source can be selected in the
a.
Use the  and  keys to adjust the control code to 010.
b.
Press the
change to the Alarm source selection sub menu.
c.
Change the selection by scrolling through these options using the  and
 keys.
The following list shows the available alarm sources:
DISP – not in use
RESULT – pressure in barg
DP_C – °C dew point
psiG – pressure in psig
DP_F – °F dew point
PPMV_I – moisture content in ppm
TOT_1 – not in use
. . .
TARE – not in use
AUX_1 – moisture content in lb/MMscf natural gas
AUX_2 – moisture content in mg/m
AUX_3 – moisture content in ppm
AUX_4 – not in use
Michell Instruments
key to confi rm this control code. The display will then
Select
R E S U L T
S O U R C E
page.
ALARM
0 1 0
S P C _ 1
ideal gas
V
natural gas
3
natural gas
V
OPERATION
43