Easidew Online User's Manual
3.3.4
Re-Transmitted Output Current Range Set-Up
The Easidew Online is provided with an
analog current loop output module which
buffers and re-transmits the current loop
input signal from the dew-point transmitter.
By default, the re-transmission output is set
as a 4-20 mA current loop (to exactly follow
the input signal, i.e. 4 mA in, 4 mA out).
For certain system processes, a 0-20 mA
current loop output may be required. The
set-up method is as follows:
Figure
26
sequence.
To change output from 4-20 mA to 0-20
mA:
1.
Press the
read
tECH
2.
Press the
fl ash between
3.
Press the key and the display will
fl ash between
4.
Press the
5.
Press the key once to display a
fl ashing 4 digit number. For the default
setting (4-20 mA) this will be 0001.
6.
Press the key once to change the
display to 0000. This selects the re-
transmission output to be 0-20 mA.
7.
Press the
value. The output current loop is now
0-20 mA. The display will fl ash between
and
out1
8.
Press the
main dew-point temperature display.
Note: The transmitter current loop output signal is now scaled at 6 mA per °C
input, while the transmitter input remains scaled at 7.5°C per mA.
Michell Instruments
shows
the
operational
key once, the display will
P
.
key and the display will
SET
and
ConF
PinP
and
out1
ConF
key to display
SET
oAt1
key to accept the new
SET
.
ConF
key once to return to the
P
key
.
.
.
Figure 26
OPERATION
P
SET
SET
Default setting
0001=4 to 20mA
New setting
0000=0 to20mA
SET
P
Main
Display
Confi gure Analog Output
31