OPERATION
3.6.2

Maximum Dew-point Value for the Fault Alarm

This is the maximum dew-point above which the fault alarm will be activated.
a.
Increase or decrease the value by pressing the  and  keys.
b.
When the desired set-point is on the display press the
confi rm. The display will then change to the next setting parameter - the
minimum pressure fault alarm set-point.
NOTE: This fault alarm set-point can only be set in °Cdp, no matter what unit
has been selected for the Main Reading Page.
3.6.3

Minimum Pressure Value for the Fault Alarm

This is the minimum pressure below which the fault alarm will be activated.
a.
Increase or decrease the value by pressing the  and  keys.
b.
When the desired set-point is on the display press the
confi rm the choice. The display will then change to the next setting
parameter - the maximum pressure fault alarm set-point.
NOTE: The unit could be either barg (B) or psig (P), depending on the choice
of pressure unit in
The value will be automatically converted if a new unit is selected.
38
2 0. 0 C
M A X _ D P
M I N _ B
– display settings sub-menu (refer to Section 3.5.1).
DISP
Promet I.S. User's Manual
Select
0. 0 8
Select
97221 Issue 5, March 2017
key to
key to