Default Values - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox

Default Values

11.
Logger:
Installation:
Sensors
Installation:
pH mode
Redox mode
A-96.250.441 / 060619
Default Values
NOTICE: The AMI pH-Redox has two different operating modes
(pH or Redox) which can be set in menu <Type of sensor>. The
instrument remains in selected operating mode even after the
<Default Values> are reset completely. Therefore this default
value list is divided in the two parts pH and Redox where
necessary.
Filter Time Const.: ..................................................................... 30 s
Hold after Cal.:......................................................................... 300 s
Relay 1 and 2 ............................................... same as in Installation
Input.............................................................. same as in Installation
Logger Interval:..................................................................... 30 min
Clear Logger:................................................................................ no
pH Sensor
Flow: Flow measurement: ....................................................... None
Parameter: Type of sensor: ......................................................... pH
Parameter: Sensor Check: .......................................................... Off
Temperature; Temp. Sensor: ....................................................... No
Default Temp. .......................................................................... 25 °C
Temp. Compensation............................................................. Nernst
Standards: pH Standard 2 ......see Chap. 4
Quality Assurance; Level: ........................................................ 0: Off
Redox Sensor
Flow: Flow measurement: ....................................................... None
Parameter: Type of sensor: ....................................................Redox
Temperature; Temp. Sensor: ....................................................... No
Temperature; Default Temp. .................................................... 25 °C
Standard ...............................................................................475 mV
Parameter:.................................................................... Meas. Value
Current loop:...................................................................... 4-20 mA
Function:.................................................................................. linear
Scaling: Range low:............................................................. 0.00 pH
Scaling: Range high: ......................................................... 14.00 pH
Scaling: Range low:................................................................. 0 mV
Scaling: Range high: ......................................................... 1000 mV
Set Buffer Values, S. 39
Set Buffer Values, S. 39
81