Maintenance; Maintenance Table - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox

Maintenance

6.
6.1.
44
Maintenance

Maintenance Table

Swansensor pH/Swansensor Redox (ORP)
Quarterly
Calibrate electrode.
Ensure buffers are not expired.
If necessary clean electrode
Yearly
Replace electrode
Swansensor pH AY/Swansensor Redox (ORP) AY
Twice per
Clean electrode.
month
Monthly
Calibrate electrode.
Ensure buffers are not expired.
If necessary clean electrode.
Swansensor pH SI or FL
Weekly
Check level in electrolyte bottle.
Monthly
If necessary, change electrolyte bottle.
Calibrate electrode.
Quarterly
Slightly open the sensor cap of the reference sen-
sor and allow a small amount of electrolyte (~5 ml)
to flow out. Fasten cap hand-tight.
Swansensor Redox (ORP) SI or FL
Weekly
Check level in electrolyte bottle.
Monthly
If necessary, change electrolyte bottle.
If necessary, correct electrode.
Quarterly
Slightly open the sensor cap of the reference sen-
sor and allow a small amount of electrolyte (~5 ml)
to flow out. Fasten cap hand-tight.