Error List - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox

Error List

7.
A-96.250.441 / 060619
Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
 Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked
 Errors which indicate a hardware failure of the instrument.
Such Errors are marked
HOLD
14:10:45
R1
7.76 pH
R2
21.7 °C
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
Messages
Pending Errors
Message List
Pending Errors
Error Code
Alarm low
<Enter> to Acknowledge
E0xx
(bold and orange)
E0xx (bold and
Error or
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and
take corrective action.
Press [ENTER].
Navigate to menu Messages.
1
Press [ENTER].
Navigate to menu Pending Errors.
1.1
Press [ENTER].
Press [ENTER] to acknowledge the
1.1.5
Pending Errors. The Error is reset and
E002
saved in the Message List.
red)
fatal Error