AMU pH-Redox
Error List
Error
E001
Alarm high
E002
Alarm low
E007
Sample Temp. high
E008
Sample Temp. low
E009
Sample Flow high
E010
Sample Flow low
E011
Temp. shorted
E012
Temp. disconnected
E013
Case Temp. high
E014
Case Temp. low
E017
Control Timeout
54
Description
Corrective action
– check process
– check programmed value, see
5.3.1.1.1, p. 71
– check process
– check programmed value, see
5.3.1.1.25, p. 72
– check process
– check programmed value, see
5.3.1.3.1, p. 72
– check process
– check programmed value, see
5.3.1.3.25, p. 72
– check sample flow
– check programmed value, see
5.3.1.2.2, p. 72
– establish sample flow
– clean instrument
– check programmed value, see
5.3.1.2.35, p. 72
– Check wiring of temperature sensor,
see
Sensor, p. 34
– Check temperature sensor
– Check wiring of temperature sensor,
see
Sensor, p. 34
– Check temperature sensor
– check case/environment temperature
– check programmed value, see
p. 72
– check case/environment temperature
– check programmed value, see
p. 73
– check control device or programming in
Installation, Relay contact, Relay 1 and
2 see
5.3.2 and 5.3.3, p. 73
5.3.1.4,
5.3.1.5,
A-96.250.441 / 060619