Swann AMU pH-Redox Operator's Manual page 28

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Installation
Install the
sensor
26
This instruction applies for both the pH and the ORP sensor.
A
KCl bottle
B
Connector cap
C
KCl supply pipe
D
Dosing tip
E
Connector
1 Unscrew and remove the blind plug [F] from the flow cell block.
2 Carefully remove the protective cap [I] from the sensor tip. Turn
it clockwise only.
3 Rinse the sensor tip with clean water.
4 Insert the sensor through the flow cell block [H] into the calibra-
tion vessel [J].
5 Tighten it hand-tight.
6 Remove the connector cap [B].
7 Screw the connector [E] onto the sensor.
8 Keep the protective caps on a secure place for later use.
9 Attach the KCl supply pipe to the dosing tip of the KCl bottle.
10 Mount the KCl bottle to the bottle holder.
A
B
C
D
E
F
G
H
I
J
F
Blind plug
G
Sensor
H
Flow cell block QV-Flow
I
Protective cap
J
Calibration vessel
A-96.250.441 / 060619