Installation - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Program List and Explanations

5 Installation

5.1 Sensors
5.1.1
5.1.1.1
5.1.2
5.1.2.1
5.1.2.2
5.1.3
5.1.3.1
5.1.3.21
A-96.250.441 / 060619
Flow
Flow measurement: Select "Q-Flow" or "None". The deltaT flow
sensor is not supported by the AMU transmitter.
Flow measurement
None
Q-Flow
deltaT
Parameters:
Type of Sensor: Set the type of installed sensor. Possible types:
Type of Sensor
pH
Redox
Sensor Check: Default setting is: off. (Only visible in pH measuring
mode).
Sensor check
off
on
The <Sensor Check> can be enabled for automatic, periodic check
of the sensor aging (measuring the impedance). The measuring
values are written in the logger file or transferred to a control room
via Profibus or Modbus. It does not indicate an Alarm.
Temperature:
Temp. Sensor: The pH measurement is temperature dependent
therefore it is possible to install a temperature sensor. Depending
on your configuration set the Temp. Sensor to:
Temp. Sensor
Yes
No
If No is set the measuring value is compensated with the default
temperature.
Default Temp.: If no temperature sensor is installed, set the default
temperature to the assumed average temperature of the sample.
The measuring value is then compensated with this value.
Q-Flow
Measuring values are displayed as pH
Measuring values are displayed in mV
65