Swann AMU pH-Redox Operator's Manual page 52

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Maintenance
Possible error
message
50
Maintenance
Calibration
Calibration
Simulation
Set Time 01.01.05 16:30:00
Enter the correct value
with the [
] or [
key.
Process Cal.
Current Value
Offset
-8.15 mV
Calibration successful
Offset error:
 Last calibration wrong.
 Electrode old or dirty.
 Cable wet or broken.
 Reference measurement wrong.
3.1
]
3.1.1.1
7.60 pH
Enter
Calibration
Process pH
Standard pH
Process Cal.
Current Value
7.78 pH
Offset
0.00 mV
Process Value
7.78 pH
Save
<Enter>
Process Cal.
Current Value
7.78 pH
Offset
-8.15 mV
Process Value
7.60 pH
Save
<Enter>
Process Cal.
Current Value
7.78 pH
Offset
-8.15 mV
Process Value
7.60 pH
Save
<Enter>
A-96.250.441 / 060619
3.1.1
3.1.1.1
3.1.1.1
3.1.1.1