Default Values - Swann AMU Powercon Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU Powercon

Default Values

10.
Operation:
Sensors
Relay Contacts
Logger
Installation:
Sensors
Signal Output 1
Signal Output 2
Alarm Relay
60
Default Values
Filter Time Const.:...................................................................... 10 s
Hold after Cal.:......................................................................... 300 s
Alarm Relay .................................................. same as in Installation
Relay 1/2 ...................................................... same as in Installation
Input.............................................................. same as in Installation
Logger Interval:..................................................................... 30 min
Clear Logger: ................................................................................ no
Flow: ........................................................................................ None
Sensor Parameters; Cell Constant ...............................0.0415 cm
Sensor Parameters; Temp. corr. ...........................................0.00 °C
Sensor Parameters; Cable length............................................ 0.0 m
Sensor Parameters; Meas. unit ............................................. µS/cm
Temp. Compensation; Comp. ...................................................none
Quality Assurance; Level ......................................................... 0: Off
Parameter: .................................................................... Conductivity
Current loop: ......................................................................4–20 mA
Function: .................................................................................. linear
Scaling: Range low: ........................................................... 0.000 µS
Scaling: Range high:................................................................ 1 mS
Parameter: ....................................................................Temperature
Current loop: ......................................................................4–20 mA
Function: .................................................................................. linear
Scaling: Range low: ...................................................................0 °C
Scaling: Range high:................................................................50 °C
Alarm Conductivity:
Alarm high:........................................................................... 300 mS
Alarm low: .......................................................................... 0.000 µS
Hysteresis: ........................................................................... 1.00 µS
Delay:........................................................................................... 5 s
Sample Flow:
Flow Alarm .................................................................................. yes
Alarm high:.............................................................................. 20 l/h
Alarm low: ................................................................................. 5 l/h
Sample Temp:
Alarm High: ............................................................................ 160 °C
Alarm Low:.................................................................................0 °C
A-96.250.331 / 230320
-1