Flow Cell M-Flow; Install Swansensor Ph/Redox Standard Or Ay - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Installation
3.5.
3.5.1
Sensors
28
To install a sensor with a shaft length of 120 mm proceed as fol-
lows:
1 Unscrew and remove the blind plug [F] from the flow cell block.
2 Carefully remove the protective cap [J] from the sensor tip. Turn
it clockwise only.
3 Rinse the sensor tip with clean water.
4 Slide the distance sleeve [C] over the sensor shaft and slightly
tighten the fixing screw.
5 Slide the union screw [D], the washer [E] and the O-ring [G]
over the sensor shaft [H].
6 Insert the sensor through the flow cell block [I] into the calibra-
tion vessel [K].
7 Tighten the union screw [D] hand-tight.
8 Remove the connector cap [A].
9 Screw the connector [B] onto the sensor.
10 Keep the protective caps in a secure place for later use.

Flow Cell M-Flow

Install Swansensor pH/Redox Standard or AY

CAUTION
Fragile parts
The pH and the ORP sensors are fragile.
 Handle with care.
 Do not spill KCl when removing the protective cap.
This instruction applies for both, the pH and the ORP sensor.
A-96.250.441 / 060619