Maintenance - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Program List and Explanations
2.5 Interface

3 Maintenance

3.1 Calibration
3.1.1
3.1.1.4
3.1.1
3.2 Simulation
3.3.1
3.3.2
3.3.3
3.3.4
3.3.5
A-96.250.441 / 060619
Only available if optional interface is installed.
Shows the programmed communication settings.
Process pH/Redox: The process calibration is based on a com-
parative measurement of the current electrode with a calibrated
comparative electrode. See
Current Value: shows the measuring value of the current
electrode.
Offset: Shows the deviation of the measuring value of the current
electrode and the calibrated comparative electrode in mV.
Process Value: Enter the measured value of the calibrated
comparative electrode.
Standard pH/Redox: Performs a standard calibration. Follow the
instruction on the screen. See
To simulate a value or a relay state, select the
 alarm relay
 relay 1 and 2
 signal output 1 and 2
with the [
] or [
Press the [Enter] key.
Change the value or state of the selected item with the [
[
] key.
The value is simulated by the relay/signal output.
Alarm Relay:
Relay 1
Relay 2:
Signal Output 1:
Signal Output 2:
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
Calibration, p.
48.
Calibration, p. 48
] key.
Active or inactive
Active or inactive
Active or inactive
Actual current in mA
Actual current in mA
] or
63