AMU pH-Redox
Product Description
ORP Electrode
Reference
electrode
Consumables
Fluidics
QV-Flow
10
For the AMU pH-Redox four types of redox (ORP) electrodes are
available.
The Swansensor redox (ORP) Standard is a combined gel
electrode for application in drinking water and swimming pools.
Gel electrodes can not be filled again and have a limited life time.
The Swansensor redox (ORP) SI is a combined electrode with
liquid electrolyte (KCl) for the measurement of redox (ORP) value
in power plants.
The Swansensor redox (ORP) AY is a combined gel electrode for
application in waste water due to additional salt supplies.
The Swansensor ORP FL for the measurement of the redox
potential in high purity water. This sensor can only be used in
combination with Swansensor Reference FL,
A-87.860.100.
Swansensor Reference FL, reference electrode for Swansensor pH
FL or Swansensor Redox FL
One 200 ml bottle of 3.5 M KCl lasts for 1 month.
The flow cell QV-Flow consists of the flow regulating valve [D], the
flow sensor [F], the flow cell block [E], the vessel [G] and a built-in
temperature sensor [B].
The sample enters at the sample inlet [C]. It flows through the flow
regulating valve [D], where the flow rate can be adjusted. Then the
sample flows via the flow sensor [F] and the flow cell block [E] into
the vessel [G], were the pH of the sample is measured. The pH val-
ue depends on the sample temperature. The measuring value of
the temperature sensor [B] is used to recalculate the pH measuring
value to the standard sample temperature of 25 °C.
The sample leaves the vessel via flow cell block through the sam-
ple outlet [H] and flows into the pressure free drain [I].
A-96.250.441 / 060619