Operation - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Program List and Explanations
3.3 Set Time

4 Operation

4.1 Sensors
4.1.1
4.1.2
4.2 Relay Contacts
4.3 Logger
4.4.1
Interval
1 s
Time
25 min
4.4.2
64
Adjust date and time.
Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
Hold after Cal: Delay permitting the instrument to stabilize again af-
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac-
tive.
Range: 0–6'000 Sec
See
Relay Contacts, p. 35
The instrument is equipped with an internal logger. The logger data
can be downloaded to a PC using the built-in RS232 interface.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, measured value uncom-
pensated, temperature, flow.
Range: 1 second to 1 hour
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
5 s
1 min
2 h
25 h
Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
5 min
10 min
5 d
10 d
30 min
1 h
31 d
62 d
A-96.250.441 / 060619