AMU pH-Redox
Product Description
2.2.2
Connections
Sample
conditions
Dimensions
14

Flow Cell M-Flow 10-3PG

Flow cell for potable water applications for the installation of three
sensors, i.e. a pH or redox sensor, a reference sensor and a tem-
perature sensor. Sensor cleaning available as an option.
33
10
10
Sample:
Cleaning water:
Sensor:
Equipped with elbow hose nozzle for 10 mm tube.
For the flow cell without electrodes!
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
Particle size:
No strong acids and bases.
No organic solvents.
Width:
Front-to-back:
Height:
Mounting:
142
90
38
9
60
G 1/4" thread
G 1/4" thread
Screw connection: PG 13.5 mm
Installation depth:120 mm
4 to 15 l/h
up to 50 °C
up to 1 bar @ 25 °C
Pressure free
below 0.5 mm
90 to 142 mm
138 mm
161 mm
3 screws M5