Connecting Sample Inlet And Outlet; Serto Fitting Stainless Steel For Qv-Flow Cell; Elbow Hose Nozzle For M-Flow Cell - Swann AMU pH-Redox Operator's Manual

Version 6.20 and higher
Table of Contents

Advertisement

AMU pH-Redox
Installation
3.3.
3.3.1
Preparation
Installation
3.3.2
24

Connecting Sample Inlet and Outlet

Serto Fitting Stainless Steel for QV-Flow Cell

Cut the tube to length and deburr it. The tube must be straight and
free from blemishes for approximately 1,5 x tube diameter from the
end.
Lubrication with lubricating oil, MoS2, Teflon etc. is recommended
for the assembly and reassembly of bigger sized unions (thread,
compression ferrule).
1 Screw on the union nut by hand until finger tight. At the same
time, push the tube against the body.
2 Tighten down the union nut 1¾ rotation using an open ended
spanner. Hold Body from turning with a second wrench.
A
B
C

Elbow Hose Nozzle for M-Flow Cell

Use plastic tube (FEP, PA, or PE 10 x 12 mm) to connect the sam-
ple inlet and outlet.
A
A
Body
B
Compression ferrule
C
Union nut
D
Tube
D
A
Plastic tube 10 x 12
B
Elbow hose nozzle
B
A-96.250.441 / 060619