AMU pH-Redox
Maintenance
6.5.
52
Longer Stop of Operation
1 Stop sample flow.
2 Shut off power of the instrument.
3 Unscrew and remove the connectors from the electrodes.
4 Put the connector caps on it.
5 Remove the electrodes from the flow cell.
6 If available remove the KCL bottle form the bottle holder.
7 Rinse the electrodes well with clean water.
8 If available remove the KCl supply pipe from the KCl bottle and
close the supply pipe with a plug.
9 If available dispose the KCl according to your local regulations.
10 Fill 3.5 molar KCl (if not available: clean water) into the protec-
tive caps and put them on the tips of the electrodes.
11 Store the electrodes with the tips pointing downwards in a frost-
protected room.
12 Empty and dry the calibration vessel
CAUTION
Damage of pH or ORP sensor
Wrong storage will damage the pH or ORP sensor.
Never store the sensors dry.
Store the sensors with tip pointing downwards in a frost-pro-
tected room.