:COMPliance:USB:TEST
Command Syntax
:COMPliance:USB:TEST <test>
<test> ::= {DHSS | HHSS | DLSS | HLSS | DFSS | HFSS}
The :COMPliance:USB:TEST command selects the type of signal quality test to
perform:
• DHSS — Device Hi-Speed Signal Quality.
• HHSS — Host Hi-Speed Signal Quality.
• DLSS — Device Low Speed Signal Quality.
• HLSS — Host Low Speed Signal Quality.
• DFSS — Device Full Speed Signal Quality.
• HFSS — Host Full Speed Signal Quality.
For Hub upstream testing, select the Device tests.
For Hub downstream testing, select the Host tests.
Low and Full Speed tests require two or three single-ended probes.
Hi-Speed tests are available only on 1.5 GHz bandwidth oscilloscopes, and a
differential probe or SMA cable connections are required.
Note that the N2750A Series InfiniiMode differential probes cannot be used with
the Hi-Speed Signal Quality test fixtures because they have damping resistors and
the USB test fixture also has damping resistors specifically designed for the
E2678A socketed probe heads.
Query Syntax
:COMPliance:USB:TEST?
The :COMPliance:USB:TEST? query returns the selected signal quality test type.
Return Format
<test><NL>
<test> ::= {DHSS | HHSS | DLSS | HLSS | DFSS | HFSS}
See Also
•
•
•
•
•
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1610)
":COMPliance:USB:TEST:TYPE"
":COMPliance:USB:SOURce:DPLus"
":COMPliance:USB:SOURce:DMINus"
":COMPliance:USB:SOURce:ADJacent"
":COMPliance:USB:SOURce:DIFFerential"
on page 377
on page 374
on page 373
on page 371
on page 372
12
:COMPliance Commands
375