Chapter 1. Programming Site Data

1.9.3 Option Card Alarms

1. Use the  and [ and the
2. Use the
keys to set up alarm relays A, B, or C, respectively.
[F1]-[F3]
Note: The set up of alarm A is used as an example. Identical procedures would be used to set up the other alarms.
= OFF to disable Alarm A and return to the previous prompt, or press
3. Press
[F1]
FAULT to specify the type for Alarm A.
4. Press [F1] = NO for standard operation or [F2] = YES for Failsafe operation.
5. Do one of the following:
If you selected
FAULT
If you selected
HIGH
a. Use the  and [ and
Figure 12 on page 88.
b. Enter a value for the trigger point of the alarm and press
6. To select the Fault Type that will trigger the alarm, press
7. Press
[F1]
-
[F3]
to set up another alarm or press
For instructions on setting up additional inputs and/or outputs proceed to the appropriate section. Otherwise, press
to return to the main
[EXIT]
16
[F1]-[F4]
keys to select the desired slot number.
, proceed to step 6.
, complete the steps below.
or
LOW
keys to specify the desired Output Measurement parameter, as shown in
[F1]-[F4]
menu prompt.
I/O
.
[ENT]
F1 = FLOW, F2 = OTHER
[EXIT]
to select another slot for set up.
DigitalFlow™ GF868 Programming Manual (1-Channel)
[F2] = HIGH, [F3] = LOW or [F4]
(non-flow), or
.
F3 = ALL
=