5 Installation and startup
5.4 Commissioning
24
6. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see Switching the unit ON and OFF [ → 25]).
9. Switch the pump on (press the
10. Fill the water tank up again until the filter insert is completely
immersed (up to the bottom edge of the cap thread).
5.4.8.2

CEREC MC XL with premium package

inLab water tank
A
Water tank
A
Filter inserts
B
Tank cap
The water tank has been drained, see "Removing water from the
Fragment Filling the inLab water tank
unit" [ → 58].
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off.
NOTICE
Damage to surfaces!
When undiluted, DENTATEC grinding additive etches plastic surfaces
and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter inserts are completely immersed
(up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait briefly until the filter inserts are soaked full and add a
corresponding amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand.
Do not use the tank cap opener for this.
Operating Instructions CEREC MC XL
"Pump"
button) to fill the water circuit.
B
A
D
C
Tank
D
Tank drain
Sirona Dental Systems GmbH
C
61 46 885 D 3439
D 3439.201.02.19.02 12.2014