3.7.2 ITC5xx operation error messages
[1301, "Interlock is open"]
Category:
Possible reason:
(Refer to chapter 2.22, "Using the interlock input" on page 52)
[1302, "Open circuit"]
Category:
Possible reason:
[1303, "Over temperature"]
Category:
Possible reason:
[1304, "Internal power failure"]
Category:
Possible reason:
[1306, "No calibrating of PD during laser on"]
Category:
Possible reason:
[1307, "No setting of ILD during constant power mode"]
Category:
Possible reason:
Execution Error
Try to switch on the output while the Interlock line is open.
Execution Error
Try to switch on the output while the able to the laser diode has
opened or the compliance voltage is not high enough.
Execution Error
Try to switch on the output while the internal temperature is too
high. Wait until the ITC5xx has cooled down. Maintain proper air
flow.
Execution Error
Try to switch on the output while an internal power failure
occurred.
Execution Error
The monitor diode cannot be calibrated during the laser diode
output is switched on.
Execution Error
The set value of the laser diode current can not be changed
during constant power mode.
3.7 Error messages of the ITC5xx