3.6 ITC5xx specific commands
3.6.15 Temperature Sensor (SENS)
Programming:
Syntax:
":SENS AD"
":SENS THL"
":SENS THH"
Assumption: The TEC output is switched off.
Description:
Selects the type of temperature sensor. Sets the programmed
temperature / resistance to default values.
Reading:
Syntax:
":SENS?"
Response:
[:SENS AD]
[:SENS THL]
[:SENS THH]
Description:
Queries the type of the temperature sensor.
3.6.16 Switching the TEC on and off (TEC)
Programming:
Syntax:
":TEC OFF"
":TEC ON"
Default Value: OFF
Assumption: To switch the TEC on there must be no device errors (no sensor, TEC
open, over temperature, ...).
Description:
Switches the TEC output on or off.
Reading:
Syntax:
":TEC?"
Response:
[:TEC OFF]
[:TEC ON]
Description:
Queries the state of the TEC output.
ITC500 / page 80
(AD590 family)
(thermistor, 20 kΩ range)
(thermistor, 200 kΩ range)