6.1 Weight Calibration

The values of the footscan device are proportionate to the pressure on each sensor,
up to a factor. This weight calibration factor is identical for all sensors but might vary in
time due to wear. Footscan 9 relies on an up-to-date value of the weight calibration
factor to guarantee accurate measurements. Weight calibration recomputes the
calibration factor for each device.
Weight calibration uses a measurement of a subject with known weight. Based on such
measurement with known characteristics footscan 9 is able to determine the weight
calibration factor accurately.
rs scan recommends performing weight calibrations regularly. Set the desired time
interval between calibrations in the Settings dialog (Section 6.4).
(a)
Set weight and shoe
size.
Figure 16:
The weight calibration dialog (Figure 16a) needs the weight and shoe size of the
subject performing the calibration measurement. Click
calibration dialog. Click
Device lists all devices (Figure 16b). Devices marked with a
The dialog displays the Calibration factor and
Click
Start calibration to start the weight calibration for the selected device. The
preview shows the activity on the plate. Walk over the plate to perform a dynamic
measurement and click
Click
Close to close the weight calibration dialog.
The weight calibration needs to recognize at least one foot to succeed. The
accuracy of the weight calibration increases with the number of recognized
feet. Generally 6-8 feet suffice to obtain a reliable calibration factor.
rs scan recommends to send the footscan system to rs scan yearly for
replacement of the Pressure Sensitive Layer (PSL) and calibration. Contact
the rs scan support department at
(Return Material Authorization) procedure.
RS-PRD-00562-016
®
®
(b)
The weight calibration dialog uses a dynamic measurement with known
characteristics to calibrate each device properly.
Next to proceed.
Finish. The dialog updates Calibration factor and
®
Perform a dynamic measurement to calibrate the
device(s).
Exit to leave the weight
Date of the selected device.
®
support@rsscan.com
need weight calibration.
Date.
to obtain our RMA
64

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the footscan and is the answer not in the manual?

Table of Contents

Save PDF