3.1 Weight Calibration
Weight calibration uses a measurement of a subject with known weight to scale the
pressure data appropriately. rs scan recommends performing a weight calibration
regularly to assure accurate measuring. Refer to Section
Fill in the Weight and
size of the measurement subject.
These
are
Choose
Next to continue.
The Calibration wizard shows the
Calibration factor and
the last weight calibration of the
selected
Device. Click
calibration.
enabled and shows the activity on
the plate. Walk over the plate 6-8
times in the gait direction (Section
2.4). The wizard estimates a
calibration factor based on the
feet detected on the device. Click
Finish
calibration.
After a successful calibration the
wizard shows the updated
Calibration factor
C l i c k
Calibration wizard and to return to
the Home screen.
RS-PRD-00562-016
Shoe
mandatory
fields.
Date of
Start
The
Preview
is
to
complete
the
and
Date.
Close to exit
the
6.1
for more information.
41