Removal Of Option Exm, External Mixing Parts From Original Front Frame - Keysight Technologies N9030A Installation Note

Pxa signal analyzer. option mtu, multitouch user interface upgrade kit option mtp, multitouch user interface upgrade kit with pc6 cpu
Hide thumbs Also See for N9030A:
Table of Contents

Advertisement

Removal of Option EXM, External Mixing Parts from Original Front Frame

1. Refer to
Figure
20. Disconnect semi-rigid coax cable W29 from the Ext Mixer connector using a
5/16 inch open-end wrench. W29 will be re-used.
Figure 20
Disconnect W29 from Ext Mixer Connector
Installation Note N9030-90098
Option MTU and MTP, MultiTouch User Interface Upgrade Kits
This procedure applies only to instruments with Option EXM, External Mixing
33