Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Backup Calibration Data
1. With the instrument turned on and the signal analyzer application running as normal, connect
a USB mouse to one of the instrument USB ports.
2. Locate a USB storage device and insert it into one of the available instrument USB ports.
(Perform a virus scan on this device before use)
3. On the front panel of the instrument, press System, Alignments, Backup or Restore Align
Data.... Click on OK. The signal analyzer application will close and the Alignment Data Wizard
will start. This might take a few minutes.
4. The Alignment Data Wizard window as shown in
Figure 1
Alignment Data Wizard
5. Follow the on-screen instruction to Backup the calibration data to the USB memory device.
10
Figure 1
will appear.
Installation Note N9030-90098