Option MTU and MTP, MultiTouch User Interface Upgrade Kits
Figure 35
W29 and W17 Properly Aligned
7. Refer to
Figure
36. Use a ¼" open-end wrench to prevent the SMA female connector on W17
from twisting. Use a 5/16" torque wrench to torque the nut on W29 to 10 inch-pounds.
Figure 36
Torque Cable W17 onto W29
8. If Option EXM, External Mixing, is not installed and the Keypad Overlay which accommodates
Opt EXM (N9020-80165) is installed, locate the black hole plug (6960-0076) in the kit. Insert
the hole plug in the EXT MIXER port opening in the upper right corner of the front panel.