Option MTU and MTP, MultiTouch User Interface Upgrade Kits
3. Refer to
Figure
15. Pull the front frame carefully away from the chassis. Remove the ribbon
cable W1 from the A8 Motherboard.
Figure 15
Front Panel Cable
4. If the instrument has Option BBA (BBIQ inputs), there will be an additional cable (W24) that
will need to be removed. Use a ¼" nut driver or open end wrench to loosen and remove the
screws securing W24 to the A18 BBIQ Interface Board. Remove the two 0515-0372 screws
securing the core bracket to the display shield.
5. Pull the Front Frame Assembly carefully away from the chassis. Remove the ribbon cable W1
from the motherboard.
28
W1 may have locking springs on each side. Depress the spring on each
side of the connector to remove from the motherboard.
Installation Note N9030-90098