Condumax II User's Manual
D.4
Variables Page 4
Variable:
INST ADDR
Adjustable Range/Options:
Description: Unique instrument address for networking. This address is used by the
MODBUS protocol to specify the location of the Condumax II instrument
in the network.
Variable:
THERMO O/S
Adjustable Range/Options:
Description: A value that is used to trim out the tolerance of the thermocouple cold
junction compensation circuit - factory setting. Do Not Adjust.
Variable:
SET DEFAULT
Adjustable Range/Options:
Description: Sets instrument to default confi guration. All the default values for all the
variables and parameters are set.
The default values are:
•
Mode
•
Signal trip point
•
Max cool time
•
Cycle time
•
Minimum cooling limit
•
mA1 o/p
•
mA1 max
•
mA1 min
•
mA2 o/p
•
mA2 max
•
mA2 min
•
Wdp alarm s/p
•
HCdp alarm s/p
•
Deg C or F
•
Pressure
•
Internal temperature set-point
•
Heating temperature
•
Heat type
•
Heat ramp
•
Differential cooling limit
•
Water dew-point channel
Michell Instruments
0-31
-10 to +10°C
NONE
Measure
275 mV
4 minutes
10 minutes
-35°C (-31°F)
HCdp
+50°C (+122°F)
-50°C (-58°F)
Wdp
+50°C (+122°F)
-50°C (-58°F)
0°C (+32°F)
0°C (+32°F)
°C
psig
+20°C (+68°F)
+50°C (+122°F)
Absolute
3 minutes
+60°C (+140°F)
Dew point
APPENDIX D
63