APPENDIX D
Variable:
CYCLE TIME
Adjustable Range/Options:
Description: The
Variable:
MIN COOL
Adjustable Range/Options:
Description:
MIN COOL
Variable:
RESET LOG
Adjustable Range/Options:
Description: Resets the logging statistics so that the variation in measurements can be
60
10 to 60 minutes in 5-minute steps (default: 10 minutes)
is the frequency at which measurement cycles are
CYCLE TIME
repeated.
-100 to +100°C (default –35°C)
(Minimum Cooling Limit) should be set to 55°C lower than the
analyzer Main Unit internal operating temperature / 50°C below the air
temperature inside the sampling system enclosure.
NOTE: If the
MIN COOL
cooling cycle without detection of HCdp (
measurement cooling cycle is aborted and the recovery cycle commences.
On such occurrences, the analyzer measurement display and outputs for
HCdp will update to the temperature achieved during the cooling cycle and
the '
HCdp below cooling limit
of the display. In such cases the Condumax II confi rms that the HCdp is
at an exceptionally low level, lower than the
measurement of HCdp will recommence automatically at the next cycle
when the value rises above the
The same measurement cooling cycle abort/automatic resume function
applies to other relevant program settings:
triggering '
HCdp below cooling limit
NONE
recorded from a user defi ned start point.
Condumax II User's Manual
temperature is reached during a measurement
TRIP POINT
' message will be displayed at the bottom
MIN COOL
setting.
MIN COOL
MAX COOL
' message.
97081 Issue 28.1, July 2018
not reached), then the
setting. Normal
and
,
DIFF COOL