Section 4
Service
Replace the Interior
4-4
Model SHKE8000 Series Shakers
1. Locate the interior light on the ceiling of the chamber.
Light
2. To access the bulb, remove the cover from the light by turning it
counterclockwise. Set this aside.
3. Pull the bulb straight out and discard.
Caution Do not touch the new bulb with bare fingers. Oil from your skin
may damage bulb integrity.
4. Carefully align the two pins
of the new bulb with the
holes in the socket (Figure
4-6). Seat the bulb securely.
5. Install the cover onto the
light by turning it clockwise.
Make sure the cover is
finger-tight.
Fuses
Two fuses are located on the front of the relay enclosure located in the
lower right area of the stackable cabinet. Gain access to the relay enclosure
by removing the kick panel on the lower front of the cabinet. It is held in
place by a screw on each end of the panel and two tabs at the top. Spare
fuses are taped to the inside of the kick panel. To access the fuses, remove
the retainer/cover plate holding the relay enclosure in place. See Figure 4-7.
Figure 4-7. Relay Enclosure with and without Cover Plate
s
Figure 4-6. Position Bulb
Thermo Scientific