Alarms and Alarm
Conditions
Thermo Scientific

Service

Section 4
Caution The procedures outlined in this section should be performed only
by qualified service personnel or people trained and certified in
electrical/mechanical repair of laboratory equipment.
With the exception of the chamber air filter, Model SHKE8000/
SHKE8000-ICE and SHKE8000-7/SHKE8000-8CE Orbital Shakers
contain no user-serviceable components. Table 4-1 lists display messages
which may help diagnose abnormal conditions.
If the microprocessor control system senses a fault, malfunction or
abnormal operating condition, alarm messages appear on the liquid crystal
display. These messages are helpful if service or repair assistance should
become necessary. Refer to the table below and to the alarm matrix on the
the last three pages of this section.
Table 4-1. Alarms
Alarm Message
Overtemp Shutdown
Undertemp Shutdown
Main Temp Sensor
Over Temp Sensor
Temperature is High
Temperature is Low
Power Failure
Cycle Complete
Check Belt
Audible is Disabled!
Platform Stalled
Check Fuse
Fault Condition
System shutdown due to over temperature condition
System shut down due to under temperature condition
Temperature sensor failed
Temperature sensor failed
Temperature tracking sensed higher temperature than setting
Temperature tracking sensed lower temperature than setting
Power failed during shaker operation
Blank screen, end of countdown cycle reached
Motor V-belt is broken or slipping
Continuously notifies operator that audible alarm is disabled
Free movement of platform is obstructed
Primary drive motor fuse is blown
Model SHKE8000 Series Shakers
s
4-1