Section 2
Operation
Set the Defrost
Temperature
2-22
Model SHKE8000 Series Shakers
If no buttons are pressed, the display automatically returns to Operating
Screen after about 15 seconds, also saving the selection to memory.
To toggle the Defrost temperature
between 12°C and 14°C, press the
down arrow, the up arrow, the
Silence button, then the down
arrow button five more times.
The screen shown at right will
appear on the display.
Press the Temp button and the screen at right will appear.
Toggle between the two Defrost temperatures using the up and down
arrows.
Press Time, Speed, or Temp to return to the previous screen or press
nothing and the display will automatically return to the Operating Screen
after about 15 seconds, also saving the setpoint to memory.
Thermo Scientific