Change Temperature

Change RPM

Change Time

Thermo Scientific
1. Press the button beneath temperature setpoint (Temp°C). The
temperature value will begin to flash.
2. Press the up or down arrows to set the new temperature in 0.1°C
increments. Hold either button to scroll.
3. Press the temperature button one more time (twice for Model
SHKE8000-7/SHKE8000-8CE) to return to the Operating Screen.
The temperature can be set over a range of 5.0°C to 60.0°C. However,
Model SHKE8000/SHKE8000-ICE may not control temperature properly
if the temperature is set less than 10.0°C above ambient temperature.
Note The heating system and temperature alarms can be disabled on
Model SHKE8000/SHKE8000-ICE by setting the temperature to 4.0°.
1. Press the button beneath the Speed setpoint. The RPM value will
begin to flash.
2. Press the up or down arrows to set the new speed in 1 RPM
increments. Hold either button to scroll.
3. Press the Speed button again to return to the Operating Screen.
The RPM can be set over a range of 25 to 400 RPM.
The Stackable Shaker manages operating time in two ways:
Hold - When Time is set to Hold, the value shown in the 'Actual' display
represents total operating time and may be reset at the operator's conven-
ience. The shaker will continue to count upwards even if the shaker door
has been repeatedly opened and closed, or turned off and on with the
power switch. The Time however, will reset to 00:00 when the Stop but-
ton is pressed, then the unit restarted by pressing the Start button.
Countdown - When the Hold setpoint is changed to Countdown, enter-
ing a time value in hours and minutes programs the shaker to operate for
that period of time and automatically shut down platform motion. The
display will show the programmed time in the Setpoint segment and the
operating time remaining in the 'Actual' display, as the microprocessor
counts down to zero. The countdown time can be set over a range of 5
minutes to 200 hours in 5 minute increments.