LDS1000
10.5.16
Related commands
10.5.17
Related commands
TM
Use
MNL
Syntax
TMnn
Parameter
nn
Description
Response
Errors
C
D
Example
TM100
KI
TB
TE
SP
XN
TN
Use
MNL
Syntax
TN
Parameter
Description
Response
TNxx,yy
Errors
D
S
Example
TN
TN5,151232
TB
TE
Controller for Laser Diode Sensors
Immediate Acquisition
RMT
ACT
1 at NBMAX. Number of measurements to be performed
(NBMAX = Max. number of measurements related to the memo-
ry installed).
Error possible
nn outside authorized interval
Immediately carries out an acquisition of nn successive measu-
rements.
The acquisition period is given by SP.
The integration constant set by KI is used.
The number of the acquisition in progress is given by XN.
XN is equal to nn when the series of acquisitions is finished.
None.
Parameter out of limit.
Execution not allowed.
| Launch the acquisition of a series of 100 measurements.
Define the integration constant.
Read error.
Read error code.
Define sample period.
Read number of measurements in memory.
Read Number of Selected Sensor
RMT
ACT
None.
Read number of sensor selected by the user.
xx is the number of the selected sensor (0 to 9).
yy is its series number.
Execution not allowed.
Communication time-out.
| Request the number of the sensor selected.
| Sensor number 5 is selected, its series number is 151232.
Read error.
Read error code.
51
Error C
EDH0170FE1010 – 09/98