Rs Reset System; Sp Define Sample Period - Newport LDS1000 User Manual

Controller for laser diode sensors
Table of Contents

Advertisement

LDS1000
10.5.09
Related commands
10.5.10
Related commands
EDH0170FE1010 – 09/98
RS
Use
MNL
Syntax
RS
Parameter
Description
Response
Error
D
Example
RS
TB
TE
SP
Use
MNL
Syntax
SPnn
Parameter
nn
Description
Response
Error
D
Example
SP50
TB
TE
TK
TM
Controller for Laser Diode Sensors
Reset System
RMT
ACT
None.
Stop work in progress and carry out a RESET on the LDS1000
electronic controller.
None.
Execution not allowed.
| Reset system.
Read error.
Read error code.
Define Sample Period
RMT
ACT
Sample period in milliseconds.
nn = 0.5 to 32000 msec, by steps of 0.5 msec.
Possible errors
nn absent
nn = 0.5 msec.
nn off limits
nn = Nearest authorized value.
Sets the sample period to nn milliseconds for TM and TK func-
tions.
None.
Execution not allowed.
| Set the sample period to 50 msec.
Read error.
Read error code.
Acquisitions with trigger on TRIGG input.
Immediate acquisitions.
46