Tt Read Measurements Saved - Newport LDS1000 User Manual

Controller for laser diode sensors
Table of Contents

Advertisement

LDS1000
10.5.21
9TT,1TP12.2,2TP258.1
Related commands
EDH0170FE1010 – 09/98
TT
Use
MNL
Syntax
xxTT
Parameter
xx
Description
Response
xxTT,1TPnn1,2TPnn2
Errors
C
D
S
Example
9TT
TB
TE
Controller for Laser Diode Sensors
Read Measurements Saved
RMT
ACT
Number of the acquisition to be proof read.
xx = 0 to MAX (MAX = value returned by XN).
Possible error
xx missing
xx = 0.
xx outside authorized interval
This command enables the measurements saved to be read
again following a TK, TM or TQ command.
If xx is a whole number other than nil within the authorized
interval, the controller returns the value of the measurement
xx. If xx = 0 or absent, the controller returns all the measu-
rements in the memory (one measurement Y, Z per line star-
ting with the number 1).
In RS-232-C mode there is a 25-msec pause between two suc-
cessive lines (response at TT or 0TT) to avoid overflow.
Reading is possible as soon as the acquisitions have begun.
Parameter out of limit.
Execution not allowed.
Communication time-out.
| Read measurement number 9.
| Measurement number 9: Y = 12.2 and Z = 258.1.
Read error.
Read error code.
54
Error C.