5. The proximity probe slope and intercept values are calculated and displayed within the wizard dia-
log.
6. Click Set to accept and calibrate the TubeSpec to these values.
If the proximity scaling calibration method is completed, both the ultrasonic sensor offset and proximity
offset calibrations must be completed consecutively.

Ultrasonic Sensor-offset Calibration Method

1. If Ultrasonic sensor offset was selected in Step 3. of To calibrate the TubeSpec, the following
Calibrate dialog appears:
2. Enter the exact diameter of the calibration rod, tighten the clamping lugnuts, pressure up the pack-
ers, and click Next.
Argus TubeSpec User Guide
NOV CTES
Chapter 6: Using OrionNET Software
61