Proximity Scaling Calibration Method

1. If Proximity scaling was selected in the previous dialog, the following 
Calibrate dialog appears:
2. Input the exact diameter of the calibration rod between the proximity probes, tighten the clamping lug-
nuts, pressure up the packers, and click Add.
3. Move the calibration rod so that a different diameter exists between the proximity probes. Tighten the
clamping lugnuts, and pressure up the packers. Measure and input this new diameter; then click Add.
4. Repeat the moving/measuring process until at least two diameter measurements are included. Click
Next to continue.