To select or clear the Purge stream for 60 seconds check
box, do the following:
(a.)
Press the down arrow key to move from the Stream
drop-down list box to the Purge stream for 60 seconds
check box.
(b.)
To select or clear the Purge stream for 60 seconds
check box, press the Select/Edit key.
(c.)
If you do not want to select a calibration type, press the
Enter key.
To select a calibration type, do the following:
(a.)
Press the down arrow key to move from the Purge
stream for 60 seconds check box to the set of
Calibration Type check boxes.
(b.)
To select Normal, press the Enter key.
(c.)
To select Forced, press the down arrow key and then
press the Enter key.
Press the Enter key to start the calibration run. See the Start
Calibration screen for information about the data displayed on
this screen.