Advertisement

System Reference Manual
7P00370-H01
5.
6.
NOTICE
One or both of the heat-pipes may stay in the base plate. Do not try to remove them
because you may bend the soft heat-pipes which will reduce their performance.
7.
6 - 34
Remove the two TCD block mounting screws (1). Do not
remove the two heat-pipe cover screws (2) because the
cover will be used to pull off the detector assembly.
Remove the detector assembly by gently pulling up with a
small rocking motion on the heat-pipe bracket.
Discard the two o-rings on the detector block.
370XA Gas Chromatograph
APRIL 2014