370XA Gas Chromatograph
APRIL 2014
Usage
The type of stream. There are four types: Analy (Analysis), Cal
(Calibration), Validate (Validation), and Unused.
Tot
The number of runs to make for each calibration or validation.
Avg
The number of most-recent calibration or validation runs to use in the
average calculation; for instance, if five calibration runs are performed
and Avg is set to 3, then the last three runs of the five will be used to
average the calibration results.
Date/Time
The time at which the first automatic calibration or validation should
be performed.
4.3.3.5

The Status Display screen

Description
The GC parameter that is being monitored. This can be changed with
MON2020's LOI Status Variables screen.
Value
The GC parameter's current value.
4.3.3.6

The Communications screen

Label
The name of the port. This can be changed using the MON2020
configuration software.
System Reference Manual
7P00370-H01
4 - 23