Set The Time; Configure The Gas Chromatograph's Ethernet Ports - Emerson Rosemount 370XA System Reference Manual

Gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

System Reference Manual
7P00370-H01
3.14.5

Set the time

1.
2.
3.
4.
5.
NOTICE
You must use MON2020 to configure daylight savings time.
3.14.6

Configure the gas chromatograph's Ethernet ports

1.
2.
3.
3 - 50
Press the enter key to go to the Main Menu.
Use the right arrow key to move to the Tools menu.
Use the down arrow key to move to the Set GC Time command
and press the enter key. The Set GC Time screen displays.
Set the current date and time.
Press the edit key to activate a field.
Press the enter key to accept an entry and to de-activate
the field.
Use the arrow keys to move to the next field.
If your country does not employ daylight savings time, clear the
Enable Day Light Savings check box, which is checked by
default.
Go to the Main Menu and select TCP/IP on the Application
menu. The TCP/IP screen displays.
Make a note of the Ethernet settings for both ports. Ethernet 1
is the RJ-45 terminal that is commonly used for local computer
access; Ethernet 2 is the port that is commonly used for
communication with a supervisory system such as a flow
computer, RTU, SCADA, or DCS.
Enter the Ethernet settings according to the network
requirements of your installation. If you intend to use Ethernet
1 for local access only, do not change the settings. Contact
370XA Gas Chromatograph
APRIL 2014