The Hardware Menu; The Heaters Screen - Emerson Rosemount 370XA System Reference Manual

Gas chromatograph
Hide thumbs Also See for Rosemount 370XA:
Table of Contents

Advertisement

370XA Gas Chromatograph
APRIL 2014
4.
5.
4.3.2

The Hardware menu

4.3.2.1

The Heaters screen

Label
The heater's name. This can be changed with MON2020.
Switch
Indicates the state of the heater:
The switch state can be changed with MON2020.
Setpoint (C)
Indicates the target temperature. The setpoint and the unit of
measurement (Celsius or Fahrenheit) can be changed with
MON2020.
Temperature (C)
Indicates the current temperature.
Current PWM
Indicates the current percentage of power being provided to the
heater.
Timed event marker
Cursor coordinates. X is time, in seconds, and Y is amplitude.
Auto: The heater is controlled by the GC.
Fixed On: The heater is controlled manually, through user
input.
Not used: The heater is shut off.
System Reference Manual
7P00370-H01
4 - 13

Hide quick links: