370XA Gas Chromatograph
APRIL 2014
Figure 4-2. The Valve Events table showing the Cal Gas Save events turned
on at 20 seconds and turned off at 200 seconds.
By default, the calibration gas saver is switched on at 20 seconds and
turned off at 200 seconds, but these values can be changed.
1.
2.
3.
4.
5.
6.
Start MON2020 and select Timed Events on the Application
menu. The Time Events window opens.
Locate the first Cal Gas Save event on the Valve Events table.
This event switches on the Cal-Gas Saver feature.
Enter a new start time in the appropriate Time field.
Locate the second Cal Gas Save event on the Valve Events
table. This event switches the Cal-Gas Saver feature off.
Enter a new end time in the appropriate Time field.
Click OK to save the changes and close the window.
System Reference Manual
7P00370-H01
4 - 41