6.6

Replace the carrier gas pressure sensor

Parts required
Tools required
1.
2.
3.
4.
6 - 30
A carrier gas pressure sensor kit (#7A00053G01).
• A 2.5mm hex or allen wrench
• An 8mm wrench
Rest the module upside down on the heater cap.
Disconnect the carrier gas pressure sensor's (1) electrical
connectors (3) from the Intelligent Module Board (2).
Use the 2.5mm hex wrench to remove the Intelligent Module
Board (IMB).
Use the 8mm wrench (or small adjustable wrench) to gently
unscrew the pressure sensor from the module base.
370XA Gas Chromatograph
APRIL 2014