System Reference Manual
7P00370-H01
run. If the analysis is not within the specifications, a Module
Validation Fail alarm will be raised, and the unit will go into Idle
mode. In this case, refer to
After the calibration cycle is completed, the 370XA will then go into
Auto mode and begin analyzing the stream gas.
4 - 44
370XA Gas Chromatograph
"Module validation fails" on page
APRIL 2014
7-14).